The Waters 996 PDA Detector (part number WAT057002) is a high-performance photodiode array (PDA) detector designed for use with Waters HPLC systems. This detector delivers simultaneous multi-wavelength detection and full spectral data acquisition, enabling precise identification, quantification, and peak purity analysis in complex chromatographic applications.
With a spectral range of 190–800 nm, the Waters 996 provides excellent sensitivity and resolution for UV-visible detection. Ideal for pharmaceutical, environmental, and biochemical laboratories, the 996 PDA integrates seamlessly with Waters Empower™ or Millennium® software and other system modules.
Key Features:
– Full-spectrum detection from 190 to 800 nm
– Simultaneous multi-wavelength monitoring
– Ideal for peak purity analysis and compound identification
– Compatible with Waters HPLC and UPLC systems
– Controlled via Waters Empower™ or Millennium® software
– Reliable and robust for routine or advanced analytical workflows
Specifications:
– Manufacturer: Waters
– Model: WAT057002
– Wavelength Range: 190–800 nm – ideal for UV-Visible spectrum analysis.
– Wavelength Accuracy: ±1 nm – ensures precise chromatographic detection.
– Linearity Range: 5% deviation over -0.1 to 2.0 AU, tested with propylparaben at 258 nm.
– Spectral Resolution: 1.2 nm per photodiode using a 512-photodiode array – for high-resolution spectral scanning.
– Baseline Noise: ≤ 1.5 x 10⁻⁵ AU peak-to-peak (dry, at 254 nm), or 5 x 10⁻⁵ AU p-p – low-noise performance for reliable quantification.
– Drift: ≤ 1 x 10⁻³ AU/hour at 254 nm – maintains signal stability over long runs.
– Operating Temperature: 4°C to 40°C (39°F to 104°F) – supports wide lab environment conditions.
– Humidity Tolerance: 20% to 80% non-condensing – ensures operational reliability in variable climates
Includes:
– Waters 996 PDA Detector WAT057002
– Power Cord
Reviews
There are no reviews yet.