The Buck 410 Mercury Analyzer is designed to measure the concentration of elemental Mercury in solution. The instrument incorporates the Hatch and Ott Cold Vapor technique with a long 400mm path length flow cell for trace analysis of Mercury in drinking , surface and saline waters, as well as domestic and industrial wastes. It provides integrated measurements in absorbance intensity, as well as sample concentration in comparison to standard solutions. The absorbance readings are determined using peak area integration.
Specifications:
– Detector: model IP28; wide range general purpose, 190-930nm
– Lenses: Supracil – amorphous silica
– Optical Filter : 25.4mm Band Pass, 253.7nm center, 26nm FHWM
– Absorbance: -0.0820 to 3.2000
– Concentration: to 5 significant digits
– Integration Period: 0.5 to 25 seconds
– Integration Type: Peak Area
– Screen Refresh : 0.5 to 1.5 seconds
– Recorder Output: 1V/ABS (-0.08 to 3.2V)
– Background Correction: In-line Deuterium Arc